題名 Banding Artifacts Along the Frequency-encoding Direction in Multi-echo Magnetic Resonance Imaging: Origin and a Remedy
作者 蔡尚岳
Chuang,Tzu-Chao ;Tsai,Shang-Yueh ;Chung,Yi-Lin ;Chou,Ming-Chung ;Juan,Chun-Jung ;Chung,Hsiao-Wen
貢獻者 應物所
關鍵詞 Magnetic resonance imaging; Multi-echo imaging; Stimulated echo; Banding artifacts
日期 2004.04
上傳時間 1-May-2014 17:54:33 (UTC+8)
摘要 We report here a highly reproducible artifact seen in standard multi-echo spin-echo imaging, in which severe banding was seen along the frequency-encoding direction for all late echo images, but not for the first echo. A theoretical analysis showed that the banding resulted from the destructive interference effects of the stimulated echo in the presence of both an inaccurate calibration of the RF pulse flip angles and a gradient area offset. Experimental results from a water phantom verified the origin. A remedy method employing crusher gradients with alternating amplitudes was proposed and implemented, which successfully eliminated the artifacts as evidenced by in vivo imaging results from human brain
關聯 J. Med. Biol. Eng., 24(2), 79-85
資料類型 article
dc.contributor 應物所en_US
dc.creator (作者) 蔡尚岳zh_TW
dc.creator (作者) Chuang,Tzu-Chao ;Tsai,Shang-Yueh ;Chung,Yi-Lin ;Chou,Ming-Chung ;Juan,Chun-Jung ;Chung,Hsiao-Wenen_US
dc.date (日期) 2004.04en_US
dc.date.accessioned 1-May-2014 17:54:33 (UTC+8)-
dc.date.available 1-May-2014 17:54:33 (UTC+8)-
dc.date.issued (上傳時間) 1-May-2014 17:54:33 (UTC+8)-
dc.identifier.uri (URI) http://nccur.lib.nccu.edu.tw/handle/140.119/65813-
dc.description.abstract (摘要) We report here a highly reproducible artifact seen in standard multi-echo spin-echo imaging, in which severe banding was seen along the frequency-encoding direction for all late echo images, but not for the first echo. A theoretical analysis showed that the banding resulted from the destructive interference effects of the stimulated echo in the presence of both an inaccurate calibration of the RF pulse flip angles and a gradient area offset. Experimental results from a water phantom verified the origin. A remedy method employing crusher gradients with alternating amplitudes was proposed and implemented, which successfully eliminated the artifacts as evidenced by in vivo imaging results from human brainen_US
dc.format.extent 1254595 bytes-
dc.format.mimetype application/pdf-
dc.language.iso en_US-
dc.relation (關聯) J. Med. Biol. Eng., 24(2), 79-85en_US
dc.subject (關鍵詞) Magnetic resonance imaging; Multi-echo imaging; Stimulated echo; Banding artifacts-
dc.title (題名) Banding Artifacts Along the Frequency-encoding Direction in Multi-echo Magnetic Resonance Imaging: Origin and a Remedyen_US
dc.type (資料類型) articleen